liquid particle counter

The instrument adopts the counting principle of light resistance (shading) designated by international Hydraulic Standards Committee, and is specially used for rapid detection of oil contamination level on site.

More details

Bookmark and Share

Instrument features

 It has the advantages of small volume, light weight, fast detection speed, high precision and good repeatability, and can work under high temperature and high pressure and other harsh conditions. Built-in micro water sensor and temperature sensor can detect water content and oil temperature together with pollution degree detection. Suitable for engine oil, gear oil, transformer oil (insulation oil), hydraulic oil, lubricating oil, synthetic oil and other oil, can be widely used in aerospace, petrochemical, transportation ports, iron and steel metallurgy, automobile manufacturing and other fields.

 

1. the principle of light resistance (shading) method, the use of high-precision laser sensor, small size, high  precision, stable performance

2. suitable for laboratory or on-site detection, also optional pressure relief device for online high pressure

measurement, real-time monitoring of particle pollution in the oil system

3.external pressure chamber can be connected to form positive/negative pressure, to achieve high viscosity sample detection and sample degassing

4. built-in data analysis system, can display the real data of the diameter of each channel and automatically determine the sample grade

5. pipe 316L and PTFE materials, meet the detection of all kinds of organic solvents and oil

6. with volume washing and time washing mode, convenient for users to use and maintain the equipment

7. Built-in ISO4406, NAS1638, SAE4059, GJB420A, GJB420B, THE PARTICLE pollution level standard  OCT17216, GB/T14039

8. built-in calibration function, according to GB/T21540, ISO4402, GB/T18854 and other standards for calibration

9. built-in data analysis system, can automatically determine the sample grade according to the standard, with

automatic data processing, printing function

10. can set any alarm level, to achieve pollution or cleanliness detection

11. built-in micro water sensor and temperature sensor

 12. Chinese and English input, one key switch, with preset, input, modify, storage functions, easy and fast operation

13. large storage, can choose to store in the instrument internal or external storage devices

14. Embedded design, high-strength shell, easy to carry, suitable for all kinds of construction machinery

 

Technical parameters

Optical source: semiconductor laser flow rate range: 20-60mL/min

 Offline test sample viscosity: ≤100cSt, pressure chamber is optional when the viscosity is high

 Online detection pressure: 0.1-0.6mpa (optional decompression device maximum pressure up to 40Mpa) Particle

size range: 1-500μm (choose different types of sensors)

Interface: USB interface, power interface

Data storage: 1000 sets of data storage space, and support usb flash drive memory sensitivity: 1μm or 4μm (C)

Limit coincidence error: 10000 granules/mL Count volume: 1-999ml

IP67Count accuracy: ±0.5 pollution level protection level: IP67

Test interval: 1-24 hours test sample temperature: 0-80℃

Water activity: 0-1AW (±

0.05AW) Water content: 0-120ppm (±10%) Operating temperature: -20-60℃

Power supply: AC 220V±10%, 50/60Hz or DC12-40V