All Categories

Contact info

Labmen instrument Technology Limited
9th floor, Block 3, District D, Liancheng Walking Street, 
Jianshe Road, Yuetang, Xiangtan, Hunan, China
Tel: 86-731-52638888
Fax: 86-731-52638889
Skype: lormanled
wechat: +86 13973220437
whatsapp:  +86 18975218081

Vapor Permeability Tester

Product Features:

For plastic films, laminated films, fabrics, leather, solar cells, building materials and other materials, as well as the vapor permeability test of tableware, health, medical, textile and leather, and other fields of materials.

More details

Bookmark and Share

Working Principle:

Weigh method. Control test environment temperature, humidity, fixed humidity difference of sample in both sides; vapor goes through the sample into the dry side. Measure the weight of sample cup according to time, and analyze the vapor permeability and vapor permeability coefficient.

Application Characteristics:

Automatic test: just set the temperature, humidity, and warm-up time and stop conditions, after that, test begins automatically; automatically determine the end of the trial and save the test results;Power-off saving function; full monitoring and automatic recording; sampling rate is adjustable;The entire testing process can reproduce; powerful graphics curve function, easy comparative testing, and analysis;Host software has a common database system, and data can be exported, for data storage, query, print, and other functions. Through the analysis of historical data, it’s easy to grasp the product historical dynamic.

Features:

1. Weighing principle: double testing methods of weight loss, weight gain, in full compliance with national standards, osmotic environment is consistent with the weighing environment, eliminate operator error

2. Dual test mode: interval weighing, continuous weighing

3. The world's top high-precision sensors (MT / Mettle - Toledo), test data is accurate, test time is short

4. Pure semiconductor temperature control, no electric wire, automatic temperature control in cooling heating, temperature control is more accurate; parallel work, higher reliability;

5. Test chamber internal environment control: biplane circulating air, automatic speed control

6. Before sample weighing, stop the wind, gas, clear the balance to ensure accurate weighing

7. Precision humidity control, various humidity control methods (drying trays, dry wind towers, humidifiers)

8. Patent no slide specimen cup, does not generate any distortion or damage to specimen, 

9. Embedded control, PC machine full process monitoring, intelligent testing process

Standards compliantASTM E96, ASTM D1653, TAPPI T464, ISO 2528, DIN 53122-1, JIS Z0208

Technical data & Introduction:

Description

Parameter  

Measuring Range

0.1 ~ 10,000g / m2 / 24h

Testing Accuracy

0.01g / m2 / 24h

Quality precision

0.0001g (Mettle - Toledo balances)

Weighing Range

210g

Drifting coefficient

1/100000 g / (10ppm g / )

Testing Time

<24 hours, when the permeability 1g / m2 / day

Sample Size

Custom sizes, the standard size of area 50cm2 or 33cm2

Sample Thickness

0-3mm, custom of greater thickness fixture

Wet Control Range

Weight loss method 0-10% RH, weight increase method 30% RH ~ 90% RH (standard 90% RH)

Wet control Accuracy

2% RH

Temperature Range

5--50

Temperature Accuracy

± 0.1

Display and Control Mode

Computer display and control

Correction Mode

External calibration

Testing Environment

Standard testing environment (23 , 1barometric pressure)

Power Supply

AC220V ± 10% 50Hz

Mainframe Size

550 (L) mm × 550 (B) mm × 450 (H) mm

Main machine Net Weight

25Kg

 

System standard configuration:

No.

Name

Quantity

Unit

Specifications

1

Test host

1

Set

LAM-11-B type, consisting of: quality sensing systems, environmental control systems, communication systems.

2

Professional Testing Software

1

Set

1 Operating System Requirements: Windows 2000 or XP or more; 2 Hardware Requirements: Memory 1G, hard disk more than 80G.

3

Permeable Cup

1

Pc

50cm2 standard permeable cup

4

Communication Cable

1

Pc

Standard RS232 communication cable, 2m long

5

Circle Sample Cutter

1

Set

63cm2 sampler suitable for mounting the sample

6

Calibration Weight

1

Pc

200g

7

sealing grease

1

Pc

Dow Corning

8

USB Converter Cable

1

Pc

USB1.1 extension cable, length 1m.

Optional

Computer

1

Set

 

Optional

Gas source

1

Set

 

8 other products in the same category: