Air permeability tester

Product function:

It is applied to various gas permeation rate detection of O2, CO2, N2 and air, etc. Also in plastic film, composite membrane, High-barrier materials, sheet, metal foil, rubber, etc. 

More details

Bookmark and Share

Operating principle:

Using the principle design of differential pressure method, put treated sample in the middle of upper and bottom chamber in advance. Pressure tightly and seal,pick up the suction to the whole system, after reaching the required vacuum degree, close bottom chamber, fill with certain test gas to high pressure chamber, and adjust upper chamber pressure, keep on two sides of the sample forming constant air pressure gradient difference, the gas is under the action of differential pressure gradient, permeating from high-pressure side (upper chamber) to vacuum side (bottom chamber).Accurately measure vacuum chamber (bottom chamber) intensity of pressure variation, calculate various barrier property parameters of test sample. Fully conform to the standard ASTM D1434.

 

Application characteristic

High speed vacuum ability, reduce the vacuum time, accelerate the degassing, reduce the test error.

Short test time, high efficiency;

Can measure toxic gases, there is no gas leak;

Automatically adjust the test pressure, fit for various pressure requirements;

Save functions when power failure;

A whole-process monitor, automatic recording, can reappear in the whole test process again

Strong graph curve function, convenient for comparison test and analysis;

Automatic testing: Set the test parameters, test automatically proceeds after starting, automatically judge if the test is over and save the test result.

There is general data base system in host software, data could be exported, could store the data, inquire, print, etc. Through the analysis of historical data, it is easy to grasp the history dynamic of the products.

 

Product feature

Pure semiconductor temperature cold and heat bidirectional control, automatically control temperature warming and cooling. Temperature range is wider and more accurate,parallel work, higher reliability.

Apply base line technology; remove the system error, test result is more accurate.

Be of accurate pressure control ability, maintaining constant pressure difference.

Adopt top grade high precision vacuum transducer, high quality component and part, precise pipeline system and pressure sensor, ensure stability and reliability of instrument running

Embedded system control, fully controlled by computer

 

Meeting Standards

Compatible with the standard of various tests:ASTM D1434ISO 2556ISO 15105-1JIS K7126-AYBB 00082003

Technical index

Technical parameter

Model

APT-61

Measurement range (cc/m2·day·0.1MPa)

0.05-50,000

test precision

0.01 cc/m2•day•0.1MPa

Test sample quantity

6

Primary sensor quantity

1

test pattern

Six cavities

Integration (mean value)

vacuum error

0.1Pa

vacuum degree

<10Pa

Vacuum efficiency

Less than 27Pa in 10 minutes.

Temperature range

5℃–55, semiconductor bidirectional temperature control (water bath type optional)

temperature control accuracy

±0.1

humidity range

595%RH(double-flowhumidity method, optional)

Humidity precision

±2%RH

test pressure

0.002MPa0.2Mpa

air source pressure

0.1MPa0.8 MPa

test gas

O2N2CO2 or other gas

carrier gas port

8

vacuum tube diameter

15

Display and control method

PC machine fully control mode

Correction mode

Base line correction, Standard film correction,vacuum transducer external correction

sample thickness

3mm

Test area

30cm2, circle

Main machine dimension

(L*B*H)

1000×500×380 mm

Weight(KG)

75Kg

Power supply

220V AC 50HZ

 

 

Standard configuration

No

Name

Quantity

Model

Unit

Remark

1

Test main machine

1

APT-61

Set

Include Vacuum sensor system,

pressure sensor,gas control system,environmental control system,communication system,display & control system, etc.

2

high-vacuum pump

1

ND051

Set

Vacuum ability:0.6Pa

3

professional test software

1

APT series

Set

operating system:Windows 2000 or above XP

hardware requirement:memory 1G,hard disk above 80G

4

communication cable

1

 

Pc

Standard RS232 communication line,2m long

5

Vacuum steel tube

1.5

15mm

meter

 

6

Gas hose

2

Diameter 8mm

meter

 

7

air valve

1

 

Pc

 

8

Circle Sample Cutter-sample

1

 

Set

30cm2sampler, cut the sample

9

Filter paper sampling plate

1

 

Set

cut filter paper

10

sealing grease

1

 

Pc

Dow Corning

11

USB conversion line

1

 

Pc

USB1.1 expansion path, 1m long

12

computer

1

 

Set

 

Optional

Gas cylinders for test(oxygen ,etc)

 