Application areas:
It is widely used in research and development, process optimization and quality monitoring in various fields such as plastics, rubber, coatings, pharmaceuticals, catalysts, inorganic materials, metal materials and composite materials.
Instrument characteristics, structural advantages
1. The heating of the furnace body adopts double-row winding of precious metal platinum and rhodium alloy wire, which reduces interference and is more resistant to high temperatures.
2. The tray sensor is made of precious metal platinum and rhodium alloy, which has the advantages of high temperature resistance, oxidation resistance, and corrosion resistance.
3. Separate the power supply and the circulating heat dissipation part from the main unit to reduce the influence of heat and vibration on the micro-thermal balance.
4. Adopt top-opening structure, easy to operate. It is difficult to move the furnace body up to place the sample, and it is easy to cause damage to the sample rod.
5. The new main engine isolates the thermal influence of the heating furnace body on the chassis and the micro-thermal balance.
6. The furnace body can be replaced according to customer requirements
Controller and software advantages
1. Imported 32bit ARM processor Cortex-M3 core, sampling speed, processing speed is faster.
2. 24bit four-channel sampling AD collects DSC signal, TG signal and temperature T signal.
3. The power supply adopts isolated power supply, no interference
4. The USB two-way communication is adopted between the software and the instrument, which fully realizes remote operation, and the parameter setting of the instrument and the operation and stop of the instrument can be carried out through the computer software.
5.7 inch full color 24bit touch screen, better man-machine interface. Temperature calibration can be achieved on the touch screen
Instrument configuration
No |
name |
QTY |
1 |
Host |
1 set |
2 |
CD |
1 pc |
3 |
power cable |
1pc |
4 |
Data line |
Two pcs |
5 |
Sample crucible |
100pcs |
6 |
10A fuse |
5pcs |
7 |
Manual |
1 pc |
8 |
Warranty |
1pc |
Technical Parameters
temperature range |
Room temperature~1550℃ |
Temperature resolution |
0.1℃ |
Temperature fluctuation |
±0.1℃ |
Heating rate |
1~80℃/min |
Temperature control method |
Heating, constant temperature, cooling |
Program control |
Multi-segment |
Balance measuring range |
1mg~3g, can be expanded to 30g |
Precision |
0.01mg |
Constant temperature time |
0~300min Arbitrarily set |
Display method |
large-screen LCD display |
atmosphere |
Inert, oxidizing, reducing, static, dynamic |
Atmosphere device |
Built-in gas flow meter, including two-way gas switching and flow rate control |
Software |
Intelligent software can automatically record TG curve for data processing and print experimental reports |
Data interface |
USB interface, dedicated software (free software upgrades from time to time) |
power supply |
AC220V 50Hz |